ESI Electronics Punch List
|
|
||||||||||||||||||||||||||
Version 1.1
|
| 1. |
Intermittent (and possibly spurious) readings of low
voltage from Renishaw quality signals from the three
stages, even when wheels are parked. Suspect that
Renishaw quality signals are OK, but that there is
a marginal connection (e.g., incompletely seated pin
in connector) to the analog board and/or Galil
controller that monitor them.
4 Jan. 2000 update: The dewar filter wheel (DWFILREN) seems to be the worst of the three in terms of these dropouts. They have gotten so bad that we have turned off the pop-up alarms in the GUI for this keyword. 31 Jan. 2000 update: It now appears that the signals to the Galils are fine. We suspect misalignment as the problem. |
        unresolved | **2. |
Several ESI ion pump controllers have failed due to the
same capacitor going bad. Need to see whether most
recent ion pump controller replacement has solved this
problem.
4 January 2000 update: So far so good. The current ion pump seems to be working well. |
        monitor | 3. | Temperature sensor readings corrupted when motors are run. |         not fixed | 4. | The wiring to the OSS temperature sensor has pulled loose and needs to be reattached. Currently it always reads -100°. |
       
scheduled         for Feb. 2000 |
6. | Repin the P2 connector on the Galil 0 analog input board and add hardware to lock the connector down. |         done | 7. | Bring out some control (power? or a black box connection?) for the shutter controller from the electronics box. |         done | 8. | Document use of the shutter controller black box. |         done | 9. | Document the procedure to change out a Galil controller. |         March |
|