After the recent NIRSPEC service, there is no sign of dewar ice absorption. The green curve is current.
Normalized Flats |
|
![]() |
![]() |
| Flat Images | Cuts |
Divided Flats |
|
![]() |
![]() |
| Divided Images | Cuts |
It's been 10 months since the NIRSPEC service. The water ice absorption is approaching 10-20%. We hope to service NIRSPEC in September which should clear this up. The current curve is purple.
Normalized Flats |
|
![]() |
![]() |
| Flat Images | Cuts |
Divided Flats |
|
![]() |
![]() |
| Divided Images | Cuts |
15 months have passed since the NIRSPEC service. NIRSPEC will be taken offline for a warm and pump service between August 14 and September 9. We are looking to stay ahead of the ice.
Normalized Flats |
|
![]() |
![]() |
| Flat Images | Cuts |
Divided Flats |
|
![]() |
![]() |
| Divided Images | Cuts |
NIRSPEC was be taken offline March 7-28 to warm the dewar and pump. This service will test whether simply warming and pumping removes ice from the dewar window, or whether the window needs to be replaced.
Normalized Flats |
|
![]() |
![]() |
| Flat Images | Cuts |
Divided Flats |
|
![]() |
![]() |
| Divided Images | Cuts |
The NIRSPEC entrance window was replaced at the end of March 2006. No icemonitor data were taken until after the earthquake-induced power outage on 15 October 2006. In post-quake icemonitor data, the ice absorption looks similar to the 6 month absorption from the previous window. Compare to 2005 Aug 19 below.
Normalized Flats |
|
![]() |
![]() |
| Flat Images | Cuts |
Divided Flats |
|
![]() |
![]() |
| Divided Images | Cuts |
After about 1 year of service, it seems the ice is returning to the NIRSPEC dewar window. Please see the images and cuts across those images below.
Normalized Flats |
|
![]() |
![]() |
| Flat Images | Cuts |
Divided Flats |
|
![]() |
![]() |
| Divided Images | Cuts |
Please click here to see the complete data set.