- Emergency
- Standard
Emergency Cryogenic Procedures
NIRSPEC is cooled continuously by use of Closed-Cycle Refrigeration
(CCR). A compressor unit in the K2 machine room provides pressurized
He gas to two cold heads on-board NIRSPEC. If this system is inoperable
for more than about 15 minutes, then NIRSPEC may require a lengthy 10 day
cooling cycle to become operable again.
- Loss of CCR Cooling: Emergency Procedures
- Loss of CCR Cooling: Vacuum Procedures
- Reading Cryogenic Temperatures
- E-mail CCR alarm
- Loss of CCR Cooling: Emergency Procedures
In the event of CCR shutdown, it is very important to assess the status
of the NIRSPEC main dewar vacuum prior to restarting the CCR compressor
unit.
- Verify CCR shutdown by inspection of unit 3; static pressure
will be high (235-270 PSI), and the green indicator light will be off.
Image of CCR Unit 3.
- Go up to NIRSPEC. Read the Edwards vacuum gauge display.
Gauge 3 (termed Lo-Vac) should be displayed.
Image of Edwards vacuum gauge display
If the vacuum is:
- Less than 1x10e-1, then it is safe to restart
the CCR compressor. Be expeditious; it would be helpful to
have an assistant waiting at the compressor to save valuable time.
- Greater than 1x10e-1, the DO NOT restart the compressor. NIRSPEC
must be vacuum pumped before cooling is restarted. Vacuum pumping
should begin as soon as possible, otherwise NIRSPEC will continue
to warm up and will require a complete cool down cycle. Read on.
- Loss of CCR Cooling: Vacuum Procedure
- Connect the pump to the NIRSPEC vacuum valve fitting.
- Start rough pumping the line. Open the NIRSPEC vacuum
valve as soon as the line is a factor of 10 below the
NIRSPEC vacuum reading.
- When the vacuum reaches 4x10e-2 (40 millitorr), start the
turbo pump. At this time the CCR compressor unit 3 can be restarted.
- Monitor the NIRSPEC vacuum. When it is less than 1x10e-4, valve
off the pump. Verify that the vacuum continues to drop with the CCR
running.
- Reading Cryogenic Temperatures
To read the nirspec cryogenic temperatures, log on to waimea
as "nspeceng" and type "cryotemps". The result is:
8:20am waimea:xnirspec> cryotemps
**************************************************
Temp 1 - Rotator Motor = 55.080000
Temp 2 - 350 2nd Stage Cold Head = 12.450000
Temp 3 - Optical Plate = 53.990000
Temp 4 - Top of Image Rotator = 76.990000
Temp 5 - LN2 Can = 54.720000
Temp 6 - 1050 Cold Head = 46.100000
Temp 7 - SCAM = 57.110000
Temp 8 - Top of TMA = 55.210000
Temp 9 - ALADDIN = 29.190000
Temp 10- Cold Strap = 31.650000
**************************************************
Note that every sensor reads below liquid nitrogen temperatures (77 K).
Example of the e-mail alarm message
This message is sent to a group of people including IS's, OA's and
facility staff when Temp 2 above exceeds 15 K. This should be an
indication that the CCR has stopped operating.
********************************************************************
ATTENTION! ATTENTION! URGENT! PLEASE PAY ATTENTION!
********************************************************************
The CCR cooling system for NIRSPEC has apparantly halted. If this
system is not restarted IMMEDIATELY, the instrument will warm up,
which means that it WILL NOT BE USABLE FOR 2 WEEKS!
It is EXTREMELY URGENT to get this system restarted. It may have
halted for the following reasons:
1. The building glycol (not instrument) system has gone down.
2. The CCR controller/compressor has lost power.
3. A leak has developed in the He lines.
********************************************************************
KEEP READING - THIS IS VERY IMPORTANT!
If the CCR compressor has been down for more than 10 minutes, then
NIRSPEC may have lost a good vacuum. Check the vacuum gauge readout
(the grey box with black face/LCD readout located on NIRSPEC's frame
near the floor; labeled "Edwards Pirani").
If the pressure is GREATER than 1 x 10^-1 Torr (100 mT),
DO NOT RESTART CCR UNIT #3 WITHOUT TOM BIDA'S APPROVAL.
********************************************************************
Inspect CCR Unit 3 in the Keck-2 Machine Room AND
Read the NIRSPEC Vacuum RIGHT AWAY!
CALL Dennis McBride (161), and/or Tom Bida (165, 883-9565) ASAP.
Standard Cryogenic Procedures
Planned cryo cycles on NIRSPEC require some coordination between
summit and HQ staff, to ensure that everything gets done in the right
order. The steps below include suggestions as to who can carry them
out most easily, but these are only suggestions. Experience shows
that the whole process is usually smoother when both summit and HQ
staff are involved.
- Procedure for a planned warmup
- Move NIRSPEC into stow if not already there.
- Lock out unwanted activity. (For Instrument Scientist)
Log on to the HQ workstation hanauma as user
nirspec and change the password using the unix command
yppasswd. This will prevent random astronomers from
attempting to start the NIRSPEC software or move motors while the
instrument is not at thermal equilibrium. Be sure to make a note of
the new password, so you can reverse the change later! Send an email
to the "instrument" alias to announce that you have done this.
- Configure the software on waimea. (IS)
Log in to the instrument host waimea as user
nirspec and perform the following steps:
- Make sure waimea is healthy. Check the time
since the last reboot with the unix command uptime. If
it's been more than a week, reboot
waimea before proceeding. This will help prevent crashes that
would interrupt temperature logging while NIRSPEC is warming up.
- Get the server running. Check the status of the
server with the recover script, and follow its instructions.
- Start temperature logging. In the home directory of
the nirspec account on waimea issue the command:
- logcryo > logs/cryotemps.warmup.[date] &
to start the temperature logging process. This will record the
temperatures on all the dewar sensors, every five minutes, in the file
you just created. For the [date] part of the filename, use
some identifier that will include the day, month and year such as
23jul01. Don't sweat it if you forget this as temps are
logged by a cron all the time.
- Turn up the Aladdin detector temperature setpoint.
To protect the Aladdin detector from condensation, it must not ever be
the coldest point in the system. The normal temperature setpoint is
30 Kelvin. Set it up to 300K with the following command:
- m change.detector.temp=300
The detector block heaters do not have enough power to take it to
300K, but setting the temperature to 300 will run the heaters flat
out, which will keep the Aladdin about 10-15 degrees K warmer than the
rest of the system.
- Silence the cron jobs.
When NIRSPEC is cold, there are cron jobs running which will send warning
emails out if they see the temperature going up. Edit the crontab and comment out the line which has nirspec_watch in it. The editor is vi.
- Send email to nirspec_info.
Send out an email describing what is going on, and that you have disabled
the cron jobs which monitor the NIRSPEC dewar temperatures.
- Touch base. (IS) Contact the summit Instrument
Technician on duty to confirm that all the steps above have been
completed, and that NIRSPEC is ready for warmup.
- Check the vacuum. (For Instrument Technician) Turn
on the Varian SenTorr vacuum gauge controller, and after it stabilizes make a
note of the vacuum reading on the "high vacuum" channel.
- Turn off red, then black switches on CCR compressor. Tag out the compressor (IT) In the Keck 2 mechanical room,
switch off the CCR compressor No. 3 that serves NIRSPEC. Immediately
advise the responsible Instrument Scientist as to the time the CCRs
were turned off, and as to the vacuum reading on the Edwards gauge
beforehand.
- Disconnect the CCR lines at the instrument
- Disconnect the CCR head power at the instrument
- Place pump station at instrument in case of need
- Monitor and Record pressures occassionally during warm up
Use this record to guide decision whether to pump or not.
- NOTE: this step should be considered only in extreme cases
where the dewar needs to be warmed up very fast, since it
involves some risk of damaging the detector and optical components within the dewar. if you have not been trained on this next step,
consult with instrument engineer for training before proceeding! Backfill
with dry nitrogen gas. (IT) First confirm that all
temperatures inside the dewar are above 100 Kelvin by either calling
the responsible Instrument Scientist, or checking the latest
temperature log file. If all temperatures are high enough, carefully
bleed dry lab-grade nitrogen gas into the dewar for just a few
minutes. Do NOT push in gas until the dewar is at ambient
atmospheric pressure! Stop while there is still some audible hissing
coming from the dewar vacuum valve. If the dewar is at ambient
pressure at 100 K, it will be substantially overpressured by the time
it reaches 273 K (typical dome ambient temp), and that is not good.
- Procedure for a planned
cooldown
- Pump for as long as practical. (IT) Before cooling down,
pump the dewar for 4-5 days if possible.
- Bake the getters. (IT) Make a note of the vacuum
reading on the pump (and dewar if possible) before beginning this
step. During the first day of pumping, connect a DC power supply to
the banana jacks on the rear module leg (by the vacuum valve) to power
the heaters in the getter material on the cold heads. The heaters are
two 25-ohm resistors wired in parallel adjust the power supply so that
the total DC power supplied to both heaters (voltage times current) is
about 12.5 watts or 6.25 watts per resistor. Make a note of the
vacuum reading on the pump (and dewar if possible) after the heaters
have been on for at least a half-hour.
- Turn off getter heaters. (IT) Run the getter
heaters as described above for only about 24 hours, then continue
pumping.
- Decide when to start CCRs. (IS) This decision will
usually be dictated by the observing schedule. It's best to allow 12
full days of CCR operation before the first science night. Subject to
that constraint, pump the dewar for as long as practical before
starting the CCRs.
- Lock out unwanted activity. (IS) If this has not
already been done as part of the warmup, log on to the HQ workstation
hanauma as user nirspec and change the password
using the unix command yppasswd. This will prevent random
astronomers from attempting to start the NIRSPEC software or move
motors while the instrument is not at thermal equilibrium. Be sure to
make a note of the new password, so you can reverse the change later!
Send an email to the "instrument" alias to announce that you have done
this.
- Configure the software on waimea. (IS)
Log in to the instrument host waimea as user
nirspec and perform the following steps:
- Make sure waimea is healthy. Check the time
since the last reboot with the unix command uptime. If
it's been more than a week, reboot
waimea before proceeding. This will help prevent crashes that
would interrupt temperature logging while NIRSPEC is warming up.
- Get the server running. Check the status of the
server with the unix command ct, and verify that you see
three server processes running. If you see none, start the server
with the command runserver. If there are 1 or 2 processes
running, or if the server won't start normally, see the troubleshooting page for help.
- Start temperature logging. If you had to restart the
server in the previous step, then any cryo logging that was running
before is now dead. Check for any lingering processes with the
command:
Kill any suspended logcryo jobs found. If there is not a working
logcryo job running, start it. In the home directory of
the nirspec account on waimea issue the command:
- logcryo > logs/cryotemps.warmup.[date] &
to start the temperature logging process. This will record the
temperatures on all the dewar sensors, every five minutes, in the file
you just created. For the [date] part of the filename, use
some identifier that will include the day, month and year such as
23jul01.
- Turn up the Aladdin detector temperature setpoint.
To protect the Aladdin detector from condensation, it must not ever be
the coldest point in the system. The normal temperature setpoint is
30 Kelvin. Set it up to 300K with the following command:
- m change.detector.temp=300
The detector block heaters do not have enough power to take it to
300K, but setting the temperature to 300 will run the heaters flat
out, which will keep the Aladdin about 10-15 degrees K warmer than the
rest of the system.
- Touch base. (IS) Contact the summit Instrument
Technician on duty to confirm that all the steps above have been
completed, and that NIRSPEC is ready for cooling down.
- Check the vacuum. (IT) Turn
on the Varian SenTorr vacuum gauge controller, and after it stabilizes make a
note of the vacuum reading on the "high vacuum" channel.
Also note the vacuum reading on the pump. Touch base with the
responsible Instrument Scientist to confirm that the vacuum is good
enough to restart CCR cooling.
- Stop pumping. (IT) Close the dewar vacuum valve,
and turn off the pump.
- Start the CCRs. (IT) In the Keck 2 mechanical room,
switch on the CCR compressor No. 3 that serves NIRSPEC. Immediately
advise the responsible Instrument Scientist as to the time the CCRs
were turned on, and as to the vacuum reading on the Varian SenTorr gauge
controller before and after.