NIRC-2 General Specifications | ![]() |
The following tables have been taken partly from the original NIRC2
proposal by Keith Matthews and Tom Soifer, and partly from more recent
information on the as-built instrument.
| Input f/ ratio | f/15 |
| Wavelength range | 0.9-5.3 microns |
| Field of view | 10x10 arcsec
(narrow camera) 20x20 arcsec (medium camera) 40x40 arcsec (wide camera) |
| Pixel scale | 0.009942 arcsec/pixel
(+/- 0.00005") 0.019829 arcsec/pixel 0.039686 arcsec/pixel |
| Filters | J, H, K, Ks, Kp, Lp, Ms, H2, Fe II |
| Pupil mask | 1 circular fixed, circumscribes the telescope pupil
1 circular rotating, inscribes the telescope pupil and masks the spiders 3 hexagonal, rotating |
| Slits | 10, 20, 30, 40, 60, 80, 120, and 160 milliarcsec |
| Coronagraphic spots | circular: 100, 150, 200, 300, 400, 600, 800, 1000, 1500, 2000 milliarcsec |
| Grisms | |
| Minimum frame time | 0.25 sec, full frame |
| Detector | 1024x1024 InSb Aladdin-3 array, 27 micron pixels
32 simultaneous outputs, grouped in 8 pixels across a row in each of 4 quadrants subarray capability 35 K operating temperature normal gain, 4 electrons/DN |
| Read noise | 60 electrons per CDS read
15 electrons for MCDS with 16 reads |
| Dark current | < 0.1 electrons/pixel/sec |
| QE | 80% at 1.7 microns |
| Well depth | ~18,000 DN at 5% linearity |