Quick Review
- Focus the telescope using MIRA: recomended
- Acquire the target
- Coarse Alignment
- Image the target field
- Image the mask: or use images of the mask acquired in the afternoon
- Identify stars
- Offset telescope to comple coarse alignment
- Fine Alignment:
- Image the mask: stars are now in the alignment boxes
- Run check_boxes/lbox for fine alignment
- Repeat Fine Alignment if necessary.
- Start Spectroscopic Observations .
Slitmask Alignment
This checklist describes the procedure for aligning
multiobject slitmasks on LRIS, assuming there are:
- alignment boxes in the multiobject slitmasks.
- pickoff mirrors for coarse alignment are NOT being used.
- LRIS IRAF routines will be used for alignment.
Focusing the telescope before aligning a slitmask is recomended.
The focus is strongly dependent on the LRIS physical
rotator position and, to a lesser extent, the telescope elevation.
It is generally recommended to re-focus the telescope when the
telescope elevation angle has changed by 30° or more,
and/or the instrument's physical rotator position
(displayed as drive on the FACSUM display) has
changed by 45° or more. In practice, this often means
that refocusing is done for each slitmask, since each mask may
have a different rotator position.
- Tell the OA the name of your target as
listed in your starlist file and also the position
angle for the instrument rotator.
- Ask to OA to run MIRA in the neighborhood
of your target.
- If you plan to observe the field for two
hours, you may ask the OA to run MIRA one hour ahead of
of your target field.
- MIRA on the blue side
should usually be run to focus the telescope because it
requires less time to re-configure the instrument.
- MIRA on the red side should
be run only if a good focus is critical to science
observations on the red side.
When the OA tells you that your target is on LRIS, check
the following information on FACSUM to verify that your
position is correct:
- Target name and coordinates
- Position angle (skypa)
- Pointing origin (should be LRIS)
In these steps, we put the stars into the alignment boxes.
Either the red or blue beam of the instrument may be used
for this step. Most observers choose to use the blue
side since it is faster to reconfigure than the red. The
following assumes the use of the blue side.
- Configure the appropriate side of the instrument
for imaging:
- Configure DS9
- Check that in ds9 Region->File Format -> XY is selected.
- Check that in ds9 Region->File Coordinate System ->
WCS is selected.
- Set slitmask to direct
- Set dichroic to any position except clear
- Set grism to clear
- Set blue filter appropriately (if you will be
taking spectra in the red, use either the
clear position or the G
filter; using U or B filter
could result in problems due to differential
atmospheric refraction)
- Set blue camera focus as appropriate for the
selected blue filter.
- Set exposure time to 10-20 seconds, as desired.
- Take an image.
- When image starts reading out,
- insert the desired slitmask
- Overlay box coordinates using the ds9
Regions->load regions
option. You may have to enter the full path name.
- Identify one alignment star on the image
and measure the pixel coordinates.
Call these x1 and y1.
- Note the positions of the corresponding box center
overlayed on the image. Call these x2 and y2.
- Move the telescope by issuing a move command.
| Display tool |
move command |
where |
| DS9 blue side |
movb x1 y1 x2 y2
| lrisserver window |
| DS9 red side |
movr x1 y1 x2 y2 |
lrisserver window |
The alignment stars should now be in their boxes;
proceed with fine alignment.
In this step, we refine the positions of the stars within
their respective alignment boxes to achieve optimal
centering of the slitmask.
- Start Guiding:
Have the OA select a guide star on the movable
guider and begin guiding. Ensure that the guide star
selected is not too close to the edge of the guider
field, since small offsets may still be required.
- Take an image.
- Start IRAF:
Select from the desktop menu:
IRAF and IDL --> IRAF --> cl + ds9
- load the keck package and the
lris subpackage
While in the IRAF xterminal, load the keck package and the
lris subpackage. This should load the routines checkboxes, do_checkboxes
lbox, do_lbox, and other routines.
- cd to the data directory.
- If you do not know the data directory, type "s outdir" in a punaluu window.
outdir = /sdata240/lris1/2000feb07 (unix path name)
- In the iraf terminal type:
cd /s/sdata240/lris1/2000feb07
The "/s" is required to access a summit directory from headquarters.
- Run do_check_boxesb or
do_lbox
- do_check_boxesb
- waits for the image to complete writing to disk
- displays the image
- reads and overlays box positions on image
- allows observers to adjust and move box positions
- launches lbox
- do_lbox
- waits for the image to complete writing to disk
- launches lbox
- Repeat steps 2 and 6 if necessary.
Until the translation and rotation are within your tolerances,
repeat steps 2 and 6. Typically, a mask is aligned when the translation
and rotation are within the measurement errors given in parenthesis.
- The lbox routine will report suggested moves in
rotation and position, as well as the
uncertainties (in parenthesis).
- Perform another alignment iteration if:
- a rotator move was required
- either the north or east offset was >
0.2 arcsec (or whatever your personal
tolerance is!)
- Reconfigure the instrument for spectroscopy:
- Set dichroic and gyrism to the appropriate
position for spectroscopy
- Set blue filter to clear position
- Set blue camera focus as appropriate for the
clear position.
- Set exposure time as desired for spectroscopy.
- Expose...phew!