Science Mosaic, Direct Mode

Last measured: 2013-Feb-04

Video Input Amplifier Gain
[e-/DN]
Readnoise
[e-]
Dark Current
[e-/pixel/hr]
1 1A 1.212±0.018 2.541±0.037 --
2 1B 1.238±0.008 2.595±0.016 4.19
3 2A 1.204±0.018 2.525±0.038 --
4 2B 1.199±0.014 2.513±0.029 3.46
5 3A 1.187±0.008 2.489±0.018 --
6 3B 1.295±0.021 2.716±0.045 4.03
7 4A 1.229±0.017 2.577±0.035 --
8 4B 1.230±0.020 2.579±0.041 3.8

Science Mosaic, Spectral Mode

Last measured: 2013-Feb-04

Video Input Amplifier Gain
[e-/DN]
Readnoise
[e-]
Dark Current
[e-/pixel/hr]
2 1B 1.238±0.016 2.595±0.034 4.19
4 2B 1.197±0.017 2.510±0.037 3.46
6 3B 1.288±0.018 2.701±0.037 4.03
8 4B 1.220±0.018 2.559±0.037 3.8
10 5B 1.173±0.020 2.460±0.042 4.71
12 6B 1.236±0.017 2.591±0.035 4.28
14 7B 1.247±0.019 2.615±0.039 3.33
16 8B 1.242±0.018 2.604±0.038 3.69

Guider

Last measured: 2008-Jul-11

Gain
[e-/DN]
Readnoise
[e-]
16.266±0.112 53.561±1.991

See Also

This page auto-generated by ./deimos_detector_data.pl