Purpose
This checklist describes the "emergency" LRIS-style procedure
for aligning multiobject slitmasks on DEIMOS when the image
header is faulty and does not contain the typical information on
the positions of the alignment boxes.
Setup
The following steps would ideally be done in the afternoon in
order to save time on sky; however, they can also be done during
the night.
- Acquire direct slitmask image. This can be done in
the afternoon via the DEIMOS Slitmask Imaging
Widget, or can be an alignment image acquired on-sky as
part of the fine alignment procedure.
- Create box coordinates file.
Alignment Procedure
This is essentially identical to the
standard slitmask alignment
procedure, with the exception of the step indicated in red
below.
- Focus telescope and complete fine alignment.
- Complete fine alignment.
In this step, you will refine the positions of the stars within
their respective alignment boxes to achieve optimal
centering of the slitmask.
- Have the OA select a guide star and begin guiding.
Ensure that the guide star selected is not too close to the
edge of the guider field, since small offsets may still be
required.
- Start a 20-40 sec direct exposure through the
slitmask.
- In your IRAF window, type
epar xbox to edit the parameters for the
xbox task. Set the input parameter
to the name of the box coordinates file you generated
earlier with check_boxes.
- In your IRAF window, type do_xbox anytime before
the image has been written to disk. The script will wait for
the current image to be written, will determine the name of the
image, and then will beep to indicate that it's time for you to
wake up and run xbox.
- To run xbox on your image:
- Set up for spectroscopy.
- Reconfigure the instrument for spectroscopy:
- Set grating to the appropriate central wavelength
(if using zeroth order imaging you can use the GO BACK button to return to
original setting).
- Set filter to the desired position for order blocking.
- Set dewar focus as appropriate for the
selected filter.
- Set the detector parameters appropriately:
- Set ReadOut to Spectral.
- Set ObsType to Object.
- Set ExpTime as desired for spectroscopy.
- Wait for FCS to settle.
Do not start your exposure until the FCS GUI shows the FCS
status as locked, or it may be smeared.
- Expose!
See Also
Go to:
DEIMOS Home Page -
Instruments Home Page -
Keck Home Page